![]()
CATER Mask
Decisions
Vaccines are not a Short Term Solution so We
Need the Safe Bubble Initiative
New Methods are Needed
to Assure a Tight Fit on Every Mask Worn
in Public
Suppliers of Disposable Masks are Focused on
Increasing Production and Not Necessarily
Holistic Solutions
Important New Findings on Performance
Differences Between Masks
N95 Tested at 98% FFE
_____________________________________________________________________________
There are many reasons why we will need to
depend on HVAC and masks as part of a Safe
Bubble Initiative (SBI) for at least several
years.
·
Hundreds of thousands of wealthy citizens of the
world will die in the next 12 months without the
SBI.
·
Millions of poorer citizens will die in the next
4 years without SBI.
·
Major impact of the vaccines will not be felt in
wealthy countries until late in the year and not
for years in poorer countries.
·
New variants of the virus are eventually likely
to evolve into one which resists the present
vaccines.
·
Prevention of future pandemics along with
efforts to reduce the impact of air pollution,
wildfires, and influenza will provide long
lasting benefits.
Mask leakage can vary from 2 to 60%.
The differences in virus transmission
are huge when you include both emitter
and recipient. At 2% escaping from the
transmitter only 2 x .02 or .04%
of the
air leaks into the recipient mask. With the 60%
leakage 36%
can be inhaled by the recipient. So 36/04
or 900 times more virus is inhaled with the
looser masks.
When you combine air leakage and media
efficiency for both transmitter and recipient
the same high ratios are found. However, other
measures such as limiting capacity are less
effective and hugely costly to the economy. You
can require a school to operate at 10% capacity
but that only provides a 10 to 1
reduction ratio.
With the Safe Bubble Initiative every entrant to
a facility would be checked to assure that his
mask is tight fitting and efficient. There would
be a tiered approach. · Tier One: Every mask type and size would be tested by the manufacturer under various motions and with individuals whose facial features match the mask size. Quantitative fit testing would be used. An approved fit testing laboratory would conduct these tests and provide the rating. The entrant to a facility would only have to show that he is wearing a rated mask and has chosen a proper size.
·
Tier One Alternative: Local operators such as
Fitness Centers or Department Stores could sell
masks and provide fit testing for each
purchaser. If they are selling one brand of mask
with five sizes, they select the appropriate
size for the purchaser and run the two minute
fit test.
·
Tier Two: Each entrant would be checked to see
that he has an accredited mask and that it is
being worn properly. This inspection could
include something as elaborate as periodic
qualitative testing or just visual inspection.
One or more visual methods need to be created. A
visual check can determine if the mask has the
following attributes:
·
Stays in position on face across a variety of
motions: walk, talk, bend over, head side to
side
·
Does not restrict field of vision
·
Adjustable noseband to seal gaps on either side
of nosebridge-mandatory
·
Head strap accessory for alternate attachment
with adjustable tension in back of head-optional
·
Trim or other design element to create a seal
between user's face and mask-optional
The manufacturer also should supply
·
Correct donning, doffing, and noseband
instructions
·
Product support to ensure correct size
·
Offer in several sizes to fit a wide range of
facial shapes and structures
·
Practical performance testing on test subjects
to determine leakage under normal activities (on
sample test subjects in lab setting)
A manual check can also be provided. Here is the
procedure recommended by Vogmask.
“Place your fingers
on the cheekbones and thumbs on sides of
chin to do inhale and exhale fitting check.
"Inhale slowly. Check to see if the facepiece
suctions slightly towards your face. No air
should leak in between your face and mask.
Additional ways to check the fit could be
developed.
Some creativity is needed. Here is one
idea.
Instead of lighting up a pumpkin why not use a
cheap disposable light to light up the openings
in the mask.
the subject would place the light in his
mask. He could manipulate it through the soft
fabric or with attached threads and change the
trajectory of the light beam to check the
periphery
while the inspector is watching.
Lights for lanterns are 30 cents each on Amazon.
So they could be given to the entrant after use.
Variations of qualitative testing could be
performed. A test only requires a hood and
method of injecting a sweet or bitter aerosol.
Many facilities have
some sort of
a line where temperature is checked.
They could be utilized with a walk
through unit. A fan filter unit and walk in
module would cost less than $5k.
Each entrant could be tested with a
modified taste test.
These are just a few approaches. There could be
much better ones if only a little time is spent
devising them. When you consider that the mask
fit is as important as vaccines and more
important than social distancing it is important
to measure it and act accordingly.
Suppliers of Disposable Masks are Focused on
Increasing Production and Not Necessarily
Holistic Solutions
The huge production requirements based on
multiple mask use per day for healthcare workers
will hamper efforts to supply enough masks and
filter media to properly address COVID.
Therefore reusable masks are a critical
necessity.
3M
replied to the McIlvaine proposed
Safe Bubble Initiative by pointing out
that all their resources are needed to make
disposable masks for healthcare workers. At the
very least there should be an effort to see how
many times disposable N95 masks can be worn by
the public before the fit and efficiency
deteriorate. Many of the two billion masks 3M
hopes to make will be needed for
applications where a disposable mask is
most appropriate. When a nurse is entering a
COVID ICU unit and then may be entering the room
of a cancer patient the risk of transmission
from one patient to another is too high to
consider wearing a previously used and
sterilized mask.
But the average individual who has probably not
been exposed to COVID during the day can wear
his mask multiple times. The N95 mask can have a
high net efficiency and even with some
deterioration will
be many times more effective than the
typical cloth
mask.
The 3M position is understandable. They are to
be commended for the great effort they have made
to increase production. It may turn out that
there really is a need for the two billion masks
they make to be used by medical personnel. If
four masks are required each day then this is
only enough for 500 million mask days or 1.7
million people for 300 days.
Here is
the 3M reply.
Thank you for your inquiry. 3M is committed to
supporting public health and the government
response to the COVID-19 outbreak. While 3M
appreciates offers to collaborate, we are unable
to commit resources to review your new product
idea at this time. Since the outbreak of
COVID-19, we have increased our production of
personal protective equipment, and we are
working to nearly double our production of
respirators, to
almost 2 billion globally, within the next 12
months. A diversion of resources would
reduce our ability to maximize production of
critical supplies and therefore reduce our
ability to have the greatest public health
impact possible.
Important New Findings on Performance
Differences Between Masks
A new study for CDC confirms many past studies.
Tight fitting and efficient masks
perform much better than the average
public mask. An abstract and summary is provided
below along with a link to the full article.
The article provides
Evaluation of Cloth Masks and Modified Procedure
Masks as Personal Protective Equipment for the
Public During the COVID-19 Pandemic
Phillip W. Clapp, PhD1,2; Emily
E. Sickbert-Bennett, PhD, MS3; James
M. Samet, PhD, MPH4; et
alJon Berntsen, PhD5; Kirby
L. Zeman, PhD2; Deverick
J. Anderson, MD, MPH6; David
J. Weber, MD, MPH3,7; William
D. Bennett, PhD2,7; for
the US Centers for Disease Control and
Prevention Epicenters Program
Key Points
Question: What
are the fitted filtration efficiencies (FFEs) of
consumer-grade masks, improvised face coverings,
and modified procedure masks commonly used
during the coronavirus disease 2019 (COVID-19)
pandemic?
Findings: In
this comparative study of face covering FFEs, we
observed that consumer-grade masks and
improvised face coverings varied widely, ranging
from 26.5% to 79.0% FFE. Modifications intended
to enhance the fit of medical procedure masks
improved FFE measurements from 38.5% (unmodified
mask) to as much as 80.2%.
Meaning:
Simple modifications can improve the fit and
filtration efficiency of medical procedure
masks; however, the practical effectiveness of
consumer-grade masks available to the public is,
in many cases, comparable with or better than
their non-N95 respirator medical mask
counterparts.
Abstract
Importance:
During the coronavirus disease 2019 (COVID-19)
pandemic, the general public has been advised to
wear masks or improvised face coverings to limit
transmission of severe acute respiratory
syndrome coronavirus 2 (SARS-CoV-2). However,
there has been considerable confusion and
disagreement regarding the degree to which masks
protect the wearer from airborne particles.
Objectives:
To evaluate the fitted filtration efficiency
(FFE) of various consumer-grade and improvised
face masks, as well as several popular
modifications of medical procedure masks that
are intended to improve mask fit or comfort.
Design, Setting, and Participants:
For this study conducted in a research
laboratory between June and August 2020, 7
consumer-grade masks and 5 medical procedure
mask modifications were fitted on an adult male
volunteer, and FFE measurements were collected
during a series of repeated movements of the
torso, head, and facial muscles as outlined by
the US Occupational Safety and Health
Administration Quantitative Fit Testing
Protocol. The consumer-grade masks tested
included (1) a 2-layer woven nylon mask with ear
loops that was tested with an optional aluminum
nose bridge and nonwoven filter insert in place,
(2) a cotton bandana folded diagonally once
(i.e., “bandit” style) or in a (3) multilayer
rectangle according to the instructions
presented by the US Surgeon General, (4) a
single-layer woven polyester/nylon mask with
ties, (5) a nonwoven polypropylene mask with
fixed ear loops, (6) a single-layer woven
polyester gaiter/neck cover balaclava bandana,
and (7) a 3-layer woven cotton mask with ear
loops. Medical procedure mask modifications
included (1) tying the mask’s ear loops and
tucking in the side pleats, (2) fastening ear
loops behind the head with 3-dimensional–printed
ear guards, (3) fastening ear loops behind the
head with a claw-type hair clip, (4) enhancing
the mask/face seal with rubber bands over the
mask, and (5) enhancing the mask/face seal with
a band of nylon hosiery over the fitted mask.
Main Outcomes and Measures: The
primary study outcome was the measured FFE of
common consumer-grade and improvised face masks,
as well as several popular modifications of
medical procedure masks.
Results: The
mean (SD) FFE of consumer grade masks tested on
1 adult male with no beard ranged from 79.0%
(4.3%) to 26.5% (10.5%), with the 2-layer woven
nylon mask having the highest FFE. Unmodified
medical procedure masks with ear loops had a
mean (SD) FFE of 38.5% (11.2%). All
modifications evaluated in this study increased
procedure mask FFE (range [SD], 60.3% [11.1%] to
80.2% [3.1%]), with a nylon hosiery sleeve
placed over the procedure mask producing the
greatest improvement.
Conclusions and Relevance: While
modifications to improve medical procedure mask
fit can enhance the filtering capability and
reduce inhalation of airborne particles, this
study demonstrates that the FFEs of
consumer-grade masks available to the public
are, in many cases, nearly equivalent to or
better than their non-N95 respirator medical
mask counterparts.
Introduction
Severe acute respiratory syndrome coronavirus 2
(SARS-CoV-2), the cause of coronavirus disease
2019 (COVID-19), is a transmissible virus that
infects the upper and lower respiratory tract,
leading to a high viral titer in saliva and
respiratory secretions. A key public health
control strategy for mitigating SARS-CoV-2
transmission is use of masks or face coverings
by the public. Masks that completely cover the
nose and mouth are effective at reducing
seasonal coronavirus and influenza transmission
when worn by infected persons and noninfected
persons who may come into contact with infected
individuals. This is supported by emerging
epidemiologic data that indicate that
community-wide use of masks can effectively
contribute to the prevention of SARS-CoV-2
transmission.
As the adoption of face coverings during the
COVID-19 pandemic becomes commonplace, there has
been a rapid expansion in the public use of
commercial, homemade, and improvised masks that
vary considerably in design, material, and
construction. Similarly, the press and social
media outlets have reported on numerous
innovative “hacks,” devices, and modifications
(enhancements) that claim to improve the
performance characteristics of conventional
masks (typically surgical or procedure masks).
Despite their widespread dissemination and use
during the pandemic, there have been few
evaluations of the efficiency of these face
coverings or mask enhancements at filtering
airborne particles. In this study, we used a
recently described methodological approach based
on the Occupational Safety and Health
Administration (OSHA) Fit Test to determine the
fitted filtration efficiency (FFE) of various
consumer-grade and improvised face masks, as
well as several popular modifications of medical
procedure masks.
Methods
Testing Procedure
Fitted filtration efficiency tests were
conducted between June and August 2020 in a
custom-built exposure chamber (US Environmental
Protection Agency Human Studies Facility in
Chapel Hill, North Carolina) as recently
described. The institutional review board at the
University of North Carolina at Chapel Hill
waived the need for study approval as well as
individual consent needed for device testing.
Briefly, a TSI 8026 Particle Generator was used
to supplement the chamber with sodium chloride (NaCl)
particles that had a count median diameter of
0.05 μm (range, 0.02-0.60 μm) as measured by a
scanning mobility particle sizer. The test
atmosphere was allowed to stabilize for 30
minutes before FFE testing. The chamber
temperature and humidity during testing ranged
from 73.4 °F to 85.1 °C and 10% to 50%,
respectively. The test atmosphere used for this
study reflects typical indoor conditions, with
exposure to small particles that are slightly
smaller than individual SARS-CoV-2 virions
(reported to range between 0.06 μm and 0.14 μm).
A sampling port was installed in each mask using
a TSI model 8025-N95 Fit Test Probe Kit to allow
sampling behind the mask. All masks were fitted
on a man (weight, 165.3 lb; height, 5 ft and
10.1 in; head size, 23.0 in) with no beard. A
pair of TSI 3775 Condensation Particle Counters
were run in single-particle analysis mode to
continuously monitor ambient particles (0.02
μm-3 μm) in the chamber just outside the face
mask and particles in the breathing space behind
the face mask at a sampling rate of 1 second.
Fitted filtration efficiency measurements were
collected during a series of repeated movements
of the torso, head, and facial muscles as
outlined by the OSHA Quantitative Fit Testing
Protocol (Modified Ambient Aerosol CNC
Quantitative Fit Testing Protocol For Filtering
Facepiece Table A–2—RESPIRATORS). The FFE
corresponds to the concentration of particles
behind the mask expressed as a percentage of the
particle concentration in the chamber air and
was measured for the duration of each test
described in the OSHA protocol (bending at the
waist, reading aloud, looking left and right,
and looking up and down). The overall percentage
of FFE is calculated as 100 × (1 − behind the
mask particle concentration / ambient particle
concentration), and the percentage of FFE and
the standard deviation were calculated across
the length of the test. The total testing time
for each mask was approximately 3 minutes.
Products Tested
Two categories of products were tested for this
study: consumer-grade face masks and medical
procedure masks with and without enhancements.
The following consumer-grade masks were tested:
(1) a 2-layer woven nylon mask (54% recycled
nylon, 43% nylon, 3% spandex) with ear loops
(Easy Masks LLC) tested with an optional
aluminum nose bridge and nonwoven filter insert
in place, (2) a cotton bandana folded diagonally
once “bandit” style or in a multilayer rectangle
according to the instructions presented by the
US Surgeon General
https://www.cdc.gov/coronavirus/2019-ncov/prevent-getting-sick/how-to-make-cloth-face-covering.html),
(3) a single-layer woven polyester/nylon mask
(80% polyester, 17% nylon, 3% spandex) with ties
(Renfro Corporation) (4) a nonwoven
polypropylene mask with fixed ear loops (Red
Devil Inc), (5) a single-layer woven gaiter/neck
cover balaclava bandana (92% polyester and 8%
spandex; MPUSA LLC), and (6) a 3-layer woven
cotton mask (100% cotton) with ear loops (Hanesbrands
Inc).
The baseline FFE of unmodified medical procedure
masks with elastic ear loops (Cardinal Health
Inc) was measured (n = 4) and compared with the
FFE of the same type of mask with various
modifications designed to enhance its function.
The following modifications were tested: (1)
enhancing the mask/face seal by tying the ear
loops and tucking in the side pleats; https://youtu.be/UANi8Cc71A0),
(2) fastening ear loops behind the head with
3-dimensional–printed ear guards; https://www.thingiverse.com/thing:4249113),
(3) fastening ear loops behind the head with a
23-mm claw-type hair clip, (4) enhancing the
mask/face seal by placing a ring of 3 ganged
rubber bands over the mask, with the center
rubber band placed over the nose and chin of the
participant and the left and right side bands
looped over each ear; “fix-the-mask” 3–rubber
band method https://www.youtube.com/watch?v=CVjGCPfRwUo),
and (5) enhancing the mask/face seal by sliding
a 10-inch segment of nylon hosiery over the
fitted mask.
Results
This study evaluated the FFE of 7 consumer-grade
masks and five procedure mask modifications. The
mean (SD) FFE of consumer-grade face masks
tested in this study ranged from 79.0% (4.3%) to
26.5% (10.5%), with the washed, 2-layer woven
nylon mask having the highest FFE and the
3-layer woven cotton mask having the lowest. The
cotton bandana folded into a multilayer
rectangle affixed to the ears with rubber bands,
as described by the US Surgeon General, provided
a mean (SD) FFE of 49.9% (5.8%). Folding the
bandana bandit style produced a similar result
(mean [SD] FFE, 49.0% [6.2%]). The tested mean
(SD) FFE of the single-layer woven polyester
gaiter/neck cover balaclava bandana was 37.8%
(5.2%). The single-layer woven polyester/nylon
mask, which is attached with tie strings, tested
at a mean (SD) FFE of 39.3% (7.2%). The nonwoven
polypropylene mask with nonelastic (fixed) ear
loops tested at a mean (SD) FFE of 28.6%
(13.9%).
As expected based on data from our previous
work, a National Institute for Occupational
Safety and Health–approved 3M 9210 N95
respirator used as a reference control provided
very high mean FFE (98.4% [0.5%]; n = 1). The
medical procedure masks with elastic ear loops
tested in this study had a mean (SD) FFE of
38.5% (11.2%), (which was lower than that of
medical surgical masks with tie strings (71.5%
[5.5%]; n = 4). Tying the ear loops and tucking
in the corners of the procedure mask to minimize
gaps in the sides of the mask increased the mean
(SD) FFE to 60.3% (11.1%). The “fix-the-mask”
3–rubber band modification and the nylon hosiery
sleeve modifications, which were also intended
to reduce gaps between the mask and the wearer’s
face, improved mean (SD) FFE to 78.2% (3.3%) and
80.2% (3.1%), respectively.
Modifications to improve the seal of the mask
against the face by increasing the tension of
the ear loops also improved FFE. Attaching the
ear loops to the ear guards device using the
center hooks (tightest option) increased
procedure mask mean (SD) FFE to 61.7% (6.5%).
Similarly, joining the ear loops behind the
wearer’s head using a claw-style hair clip
increased the procedure mask mean (SD) FFE to
64.8% (5.1%). None of the modifications tested
enhanced procedure mask FFE to the level of an
N95 respirator.
Discussion
In this study, consumer-grade masks and medical
procedure mask modifications were tested as
personal protective equipment (protection for
the wearer) against a test aerosol of 0.05-μm
NaCl particles. Although the FFE of
consumer-grade masks and face coverings was
variable, the FFE of some consumer-grade
products exceeded that of medical-grade
procedure masks. For example, the 2-layer woven
nylon mask with ear loops was tested under
various conditions, including with and without
an aluminum nose bridge, with and without a
commercially available nonwoven insert, and
after 1 wash cycle in a standard household
washing machine (air-dried on a drying rack).
The unwashed nylon mask without a nose bridge or
insert had an FFE of 44.7%. The addition of a
nose bridge reduced visible gaps around the nose
and increased FFE to 56.3%. Adding a nonwoven
filter insert to the mask with the nose bridge
in place resulted in a further increase in FFE
to 74.4%. Interestingly, the FFE of the nylon
mask (with the nose bridge but without the
filter insert) improved slightly to 79.0% after
washing. It is unclear why washing alone
improved the FFE from 56.3% to 79.0%. It may be
that the washing/drying process unraveled some
of the fibers to increase the overall filtration
surface, and thus filtration efficiency, of the
medium, or perhaps it modified the mask shape or
size in a way that improved fit, or both. The
washing/drying test was not repeated with
additional nylon masks. Further investigation to
assess the association of single and multiple
washing with mask integrity and material
disposition would be necessary to validate any
improvement in FFE.
The woven cotton mask, which comprises 3 layers
and has a thin, flexible metal nose bridge, had
the lowest FFE in this study (26.5%). The
relatively loose weave of the cotton layers,
while providing improved breathability and
comfort, may reduce filtration efficiency.
Additionally, we evaluated the FFE of improvised
face coverings, including a standard cotton
bandana and a neck gaiter balaclava bandana. The
cotton bandana, when folded either bandit style
or according to the US Surgeon General’s
instructions, achieved approximately 50% FFE,
which is better than the ear loop procedure mask
we tested. Neck gaiter balaclava bandanas have
also emerged as a popular face covering,
particularly among athletes and young adults. As
tested in this study, the single-layer gaiter,
which was made of 92% polyester and 8% spandex
and fits tightly to the wearer’s nose and mouth,
had an FFE of 37.8%. While this face covering
appeared to fit the wearer well, with no visible
gaps in the seal, it may be that the relatively
low FFE can be attributed to the low filtering
efficiency of a single thin layer of woven
material with large porosity.
For medical procedure masks, modifications that
enhanced the fit between the mask and the
wearer’s face improved FFE. Simply tying the ear
loops and tucking the corners of the mask
against the wearer’s cheeks visibly improved
mask fit and increased FFE from 38.5% to 60.3%.
The most effective modification tested was the
use of a nylon hosiery sleeve placed over the
procedure mask. This modification, which held
the mask tight to the wearer’s face, eliminated
all visible gaps and increased FFE from 38.5% to
80.2%. However, donning the nylon sleeve over
the procedure mask was cumbersome and limited
the wearer’s ability to adjust the procedure
mask. Generally, improvements in procedure mask
FFE appeared to be associated with the integrity
of the seal of the edges of the mask to the
wearer’s face, demonstrating the importance of
mask fit to maximizing filtration. While all of
the modifications described enhanced protection
against airborne particles for the wearer, not
all were comfortable or practical for extended
use. For example, the 3–rubber band
“fix-the-mask” modification created considerable
pressure on the wearer’s ears, making it
uncomfortable after only minutes of wear and
raising questions about its adoption by the
general public. While the modifications shown in
this article can improve mask fit and provide
increased filtration of airborne particles, it
is important to choose a modification in which
discomfort is not a deterrent from wearing the
mask for prolonged periods.
The full text is included in the link below to
an article in the Journal of the American
Medical Association.
N95 Tested at 98% FFE
3M™
says its Aura™ Series Particulate Respirator
9210+, N95 is a breakthrough in comfort and
convenience. This three-panel, flat-fold
disposable respirator with its innovative design
helps provide comfortable, reliable worker
protection against non-oil based particles.
The lightweight, three-panel designed disposable
N95 particulate respirator helps provide
quality, reliable, and convenient worker
respiratory protection. 3M uses a variety of
innovative technologies and features to help
meet respiratory protection and comfort needs.
3M`s proprietary filter media, 3M™ Advanced
Electret Media, filters dust and other
particles, while allowing for easy breathing.
The soft inner material provides added comfort
while the soft nose foam and adjustable nose
clip help provide a custom seal. Braided
headbands provide comfort and help minimize
pulling of hair.
Unique features to the Aura™ Series Particulate
Respirators include sculpted nose panel that
follows the contours of the nose allowing more
room for eyewear, embossed top panel that is
designed to help reduce the fogging of eyewear
from warm, moist exhaled air, and innovative
chin tab designed for ease of positioning,
donning, and adjustment. These features are
designed to enhance user comfort and help
increase wearability. The unique three-panel
flat fold design is collapse resistant and its
individual packaging allows for easy storage
prior to use. Suggested applications: Grinding,
Sanding, Sweeping, Bagging and other dusty or
arid operations. Can also be used to help reduce
inhalation of certain airborne biological
particles like mold, Bacillus anthracis,
Mycobacterium tuberculosis, etc. Example
applications include emergency or pandemic
preparedness planning, stockpiling, etc.
|